CR-IR347
Service Manual
PM - 112
009-058-00
10.20.2000 FM2732 (2)
PM - 112
10. Turning ON the High-Voltage Switch
1Y
At intervals of one year, or when 60,000 IPs have been processed
→
Turn ON.
Turn ON the high-voltage switch (S1) of the SCN08D board.
HV
S1
ON
High-voltage
switch
SCN08D board
FR7H6925.EPS
HHS label #2
ON position
11. Installing and Cleaning the Covers
1Y
At intervals of one year, or when 60,000 IPs have been processed
→
Install and clean.
(1) By reversing the procedures described in “4. Removing the Covers,” install all the
covers.
(2) Clean all the covers with a moistened cloth.
◆
CHECKS
◆
• Make sure that the screws on the covers are attached securely.
• Verify that there is no missing screw.
Summary of Contents for CR-IR 347
Page 2: ...CR IR347 CR IR347P Service Manual Preventive Maintenance PM ...
Page 22: ...CR IR347 Service Manual PM 15 3 009 058 03 08 30 2002 FM3476 PM 15 3 BLANK PAGE ...
Page 23: ...CR IR347 Service Manual PM 15 4 009 058 03 08 30 2002 FM3476 PM 15 4 BLANK PAGE ...
Page 29: ...CR IR347 Service Manual PM 21 009 058 00 10 20 2000 FM2732 2 PM 21 BLANK PAGE ...
Page 59: ...CR IR347 Service Manual PM 51 009 058 00 10 20 2000 FM2732 2 PM 51 BLANK PAGE ...
Page 73: ...CR IR347 Service Manual PM 65 009 058 00 10 20 2000 FM2732 2 PM 65 BLANK PAGE ...
Page 83: ...CR IR347 Service Manual PM 75 009 058 00 10 20 2000 FM2732 2 PM 75 BLANK PAGE ...
Page 89: ...CR IR347 Service Manual PM 81 009 058 00 10 20 2000 FM2732 2 PM 81 BLANK PAGE ...
Page 110: ...CR IR347 Service Manual PM 102 009 058 00 10 20 2000 FM2732 2 PM 102 BLANK PAGE ...
Page 119: ...CR IR347 Service Manual PM 111 009 058 00 10 20 2000 FM2732 2 PM 111 BLANK PAGE ...