CLINITEK Analyzer Operator’s Guide
Mainten
ance
6. If the white calibration bar appears dirty or discolored, perform the
following steps:
a. Wet a new cotton-tipped stick or lint-free cloth with distilled
water and gently wipe and clean the calibration bar.
b. Allow the calibration bar to air dry.
c. Inspect the surface for dust, foreign material, scratches, or
scuffs.
If you cannot completely clean the calibration bar or if the bar
has scratches, order a new test table. Contact your Siemens
representative.
d. Insert the test table and table insert, as described in step 5.
Disinfecting the Test Table and Table Insert
Disinfect the test table and the test table insert as necessary, following
your lab guidelines. Use a recommended disinfection solution for the
following reasons:
•
Prevent contamination
•
Prevent bacterial growth
•
Avoid damage to the test table and insert
CAUTION
Do not scratch the white calibration bar. Marks and stains
could cause inaccurate test results, especially for hCG tests.
Severe marks can cause errors.
Do not use solvents of any kind to clean the calibration bar.
They could destroy the bar.
CAUTION
Do not autoclave the test table or the insert because it would
destroy them.