Compare QC results to the QC manufacturer’s acceptable results list. If
the QC results are not acceptable, do not test the patient samples until
you solve the problem. Repeat QC tests until you have acceptable
results.
Quality Control Troubleshooting
If the control results fall outside the values stated in the product
instructions for use insert, try any of the following corrective actions:
•
Use a fresh strip out of a bottle or a fresh cassette out of a package
and repeat the QC test.
•
Use a fresh bottle of strips or fresh package of hCG cassettes to
repeat the quality control procedure.
•
Use a fresh control solution and repeat the quality control
procedure.
•
If you use lyophilized controls, prepare the control solution using a
fresh bottle of control product.
For more troubleshooting information, see Section 5,
,
or contact your local technical support providerfor assistance.