CLINITEK Analyzer Operator’s Guide
49
Op
erations
Completing the Cassette Full Test
Complete the testing for one cassette or continue testing one cassette
at a time until you finish testing all the cassettes you want to analyze.
To complete the cassette Full Test, perform the following steps:
1. Remove the used cassette from the test table, and dispose of it
according to your standard laboratory procedures.
2. Wipe the table insert, if necessary (see Section 4,
Weekly Cleaning of the Test Table and Test Table Insert
3. Report the results to a laboratory supervisor or physician.
4. Select
Done
to complete the test.
5. Select
Back
to return to the
Select Ready
screen.
Note
Keep the calibration bar clean to ensure accurate hCG results.
For details about cleaning the calibration bar, see