Examination
Fluoroscopy
12
/ 52
AXD3-340.620.13.01.02
Operator Manual
Dose reduction
1
In fluoroscopy with CAREVISION (pulsed fluoroscopy) or SUPERVISION the fol-
lowing dose reductions configured in the factory result related to standard fluor-
oscopy:
1
for image receptor input dose rate
Automatik-
programm
CAREVISION
15 P/s
CAREVISION
7,5 P/s
CAREVISION
3 P/s
SUPER-
VISION
1
- 50%
- 75%
- 90%
- 50%
2
- 33%
*
*
Deviating factory setting for USA
- 55%*
- 75%*
- 50%
3
- 30%
- 50%
- 65%
- 50%
Summary of Contents for AXIOM Iconos R200
Page 2: ......
Page 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Page 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Page 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Page 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...