AXIOM Iconos R200
AXD3-340.620.13.01.02
27
/ 32
System Operation
Setting the Image Geometry
Limiting the radiation field (collimation)
Rectangular and /or iris collimation
Fluoroscopy and
Digital Radiography (DR)
(optional)
You can view the collimation on the monitor under fluoroscopy. The default is iris
collimation.
The collimator is opened or closed synchronously and symmetrically to the image
receptor center.
Within the selected image intensifier format, the radiation field can be collimated
basically as desired and you can switch over from iris to rectangular collimation.
On selection of a lower ZOOM level (larger field) or image intensifier full format,
the rectangular and iris collimation are retained, or the larger field is collimated
(can be configured).
On selection of a higher ZOOM level (smaller field), the collimator always switch-
es to the smaller field (iris collimation).
Cassette exposure
The system switches over automatically to rectangular collimation on selection of
the spotfilm device or tabletop cassette exposure technique.
It is not possible to select iris collimation.
Selection / deselection
Use the key on the system remote control console for selection / deselection.
◆
Press this key
– Rectangular collimation is selected: left LED lights up.
– Iris collimation is selected: right LED lights up.
The maximum possible collimator aperture is never greater than the selected im-
age intensifier format.
The edge of the diaphragm must be visible in the fluoroscopic image.
Summary of Contents for AXIOM Iconos R200
Page 2: ......
Page 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Page 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Page 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Page 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...