System Operation
System Settings
AXIOM Iconos R200
AXD3-340.620.13.01.02
17
/ 32
Tube assembly stand
Setting the projection angle (oblique projection)
The setting range of the projection angle is maximally ± 40°.
The center of rotation of the swivel can be adjusted in height by motor drive from
10 mm to 300 mm above the tabletop to adjust it to the position of the object.
This allows isocentric viewing on the monitor.
The additional mechanical centering device ensures that the central beam is cen-
tered onto the image receptor for each projection angle.
◆
Press a key
– cranio-caudal projection (right key)
– caudo-cranial projection (left key)
If the tube assembly stand or the image receptor unit reach the end of their re-
spective travel range, the movement will stop.
To perform further oblique projections:
◆
Press the key for oblique projection again
– The tube assembly stand or receptor unit moves to a position for additional
projections
– The object centering is lost.
◆
Reposition the object by moving the tabletop longitudinally provided the range
of travel allows this.
AXIOM
For fluoroscopy, position the isocenter for oblique projection at the object height.
The object centering on the monitor will remain the same without shift if you se-
lect oblique projection.
Summary of Contents for AXIOM Iconos R200
Page 2: ......
Page 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Page 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Page 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Page 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...