AXIOM Iconos R200
AXD3-340.620.13.01.02
5
/ 32
System Operation
On-Off/Emergency Stop
Switching the system on
◆
Press the button on the generator ON/OFF console
1
.
– The entire system and all further connected devices are switched on.
– A safety test program runs automatically from this time.
– The welcome menu appears on the monitor.
The system is ready after initialization and the end of the safety test program if
at the integrated
generator control
console
❏
no error message appears in the message line of the data menu,
❏
none of the kV, mAs or ms displays flashes,
❏
a system selection is active,
on the live monitor
❏
the patient menu appears,
on the system remote
control console
❏
no displays flash,
❏
no error message appears,
on the tableside con-
trol panel
❏
no displays flash,
Emergency stop
❏
no red emergency stop button is activated.
In the case of a fault of the safety test program switch the system off and back
on within 10s. The system again starts with the automatic safety test program.
As a rule the fault is rectified by this.
If this is not the case, switch the system off. After the image system has auto-
matically switched off (monitor dark), wait 30 s and then switch it on again. In this
case the image system is run down and restarted. A possible error will be dis-
played on the monitor.
Please notify the SIEMENS Uptime Service if this is the case.
➩
Refer to the FLUOROSPOT Compact Operator Manual for further information.
Perform a functional and safety check according to the
Functional and Safety
Check
chapter.
1
If a local hardcopy camera connection is present, the associated camera and network components
should be switched on first.
Summary of Contents for AXIOM Iconos R200
Page 2: ......
Page 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Page 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Page 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Page 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...