ADVIA 1800 Operator’s Guide
Washing all the ISE lines
Materials required:
•
2 clean, empty buffer bottles
•
Probe Wash 3 solution
•
Dummy electrode
•
Phillips head screwdriver
Time: 15 minutes
Analyzer mode: Manual operation
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1.
At the Menu Panel, select
Maint
.
, then select
ISE Operation
.
2.
In the Period.wash area, select
OFF
, then select
Set
.
3.
Open the front doors and replace the buffer solution (
1
) with another buffer bottle
containing 500 mL of deionized water.
Figurer 5-36. Location of ISE Buffer bottle
4.
Loosen the thumb screw and lift the ISE cover.
5.
Disconnect the electrode connectors.
6.
Remove the thumbscrew (
1
) to release the plate that secures the electrodes and the
block containing the electrode.