142
ADVIA 1800 Operator’s Guide
a.
Return the analyzer to the READY mode.
b.
At the Operation Panel, select
Prime
.
c.
At the PRIME Settings dialog box, select
Prime 2
, then
type
10
or more in each
of the number of times fields.
d.
Select
Execute
.
Cleaning and replenishing the cuvette wash bottle
Materials required:
• Deionized water
• Cuvette detergent (wash solution)
Time: 10 minutes
Analyzer mode: STANDBY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
NOTE:
The cuvette wash bottle may be cleaned when it is refilled, but must be cleaned at
least once a month
.
1
Cuvette Wash (Detergent)
Bottle
2
Level Sensor Connector
3
Filter Cap
Figure 5-17. Cuvette wash bottle
1.
Set the analyzer to Standby.
2.
Unscrew the filter cap (
3
) at the front top of the cuvette wash bottle, then pull up the
tube with the filter.
3.
Disconnect the cuvette wash bottle level sensor connector (
2
), then turn it counter-
clockwise and pull it out.
WARNING
Before handling any reagents read the warnings on page 196.
CAUTION
Make a note of the bottle position on the shelf to avoid mixing up the fluid
bottles.
Summary of Contents for ADVIA 1800
Page 102: ...102 ADVIA 1800 Operator s Guide ...
Page 108: ...108 ADVIA 1800 Operator s Guide ...
Page 182: ...182 ADVIA 1800 Operator s Guide ...
Page 188: ...188 ADVIA 1800 Operator s Guide ...
Page 194: ...194 ADVIA 1800 Operator s Guide ...
Page 198: ...198 ADVIA 1800 Operator s Guide ...
Page 204: ...204 ADVIA 1800 Operator s Guide ...
Page 212: ...212 ADVIA 1800 Operator s Guide ...