Maintenance
115
Daily maintenance
Inspecting and cleaning the probes
Materials required:
• Phillips screwdriver
• Alcohol prep pad or lint-free towels and 5% bleach
solution
Time: 10 minutes
Analyzer mode: STANDBY
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
1.
Visually inspect each probe daily.
1
Dilution Probe (DPP)
2
Sample Probe (SPP)
3
Reagent Probe 1 (RPP1)
4
Reagent Probe 2 (RPP2)
Figure 5-1. Location of probes
•
Replace any clogged probes. See Replacing the Probes in the online Operator’s
Guide.
•
Clean all probes daily (proceed to step 3).
TIP
Perform the Shutdown Wash and Weekly Wash as scheduled, to prevent the probes
from clogging.
2.
Set the analyzer to Standby.