8 Physiol ogic Function
8 - 6
Instructions for Use
To
Do this
Display or hide a trace
on an image
Select
ECG
.
Enable trigger markers
at specific points in the
cardiac cycle
1. Enable the first ECG trigger.
●
Press the first
Delay
control.
○
For systems without a touch screen, select
T1
on the image menu.
2. Enable the second ECG trigger.
●
Press the second
Delay
control.
○
For systems without a touch screen, select
T2
on the image menu.
The system displays a trigger as a blue vertical line on each cycle of the ECG
trace. The first trigger is twice the length of the second trigger.
Each trigger defines the time when the system updates (refreshes) the 2D-mode
image. If multiple 2D-mode images are displayed, the system updates only the
real-time image. For example, if Dual format is selected with only the left image
displayed in real-time, the system updates only the left image.
Note:
If you activate a mixed imaging mode that includes M-mode or Doppler,
the system deactivates the trigger function.
Set the trigger delay(s)
1. Set a trigger delay for Trigger 1.
●
Rotate the first
Delay
control.
○
For systems without a touch screen, select
Delay
and then rotate [
Menu
].
2. Set a trigger delay for Trigger 2.
●
Rotate the second
Delay
control.
○
For systems without a touch screen, select
Delay
and then rotate [
Menu
].
Each trigger delay setting determines the position of the related trigger. The system
retains the trigger delay setting(s) until the system is powered off.
Note:
Increasing the trigger delay positions the trigger farther to the right of the
QRS complex.
ECG Troubleshooting Guide
Symptom
Possible Cause
ECG trace is not present, or
displays as a flat line
The ECG cable is not properly attached to the system.
All three leads are not connected to the patient.
Leads are poorly connected to the cable.
Leads are poorly connected to the patient.
Leads are incorrectly placed on the patient. Remove the ECG patch, clean the skin,
and attach a new ECG patch to the patient.
Gain is set too low.
Cannot position a trigger
No R-wave present.
Too much movement by the patient.
ECG trace is noisy, ragged,
or erratic
A lead or leads may be detecting muscle movement. Inspect the ECG patch
placement and reposition the lead(s) on the patient as necessary.
The conductive gel pad of an ECG patch has evaporated (dried out). Replace the
ECG patch with a new one (check the expiration date of the new patch).
Trigger is not visible on the
trace
A trigger may be superimposed on the R-wave.
Summary of Contents for Acuson S2000
Page 12: ...1 Introduction 1 2 Instructions for Use ...
Page 14: ...1 Introduction 1 4 Instructions for Use System Review Example of the ultrasound system ...
Page 84: ...2 Safety and Care 2 54 Instructions for Use ...
Page 86: ...3 System Setup 3 2 Instructions for Use ...
Page 112: ...3 System Setup 3 28 Instructions for Use ...
Page 114: ...4 Examination Fundamentals 4 2 Instructions for Use ...
Page 144: ...5 Transducer Accessories and Biopsy 5 2 Instructions for Use ...
Page 196: ...7 Specialty Transducers 7 2 Instructions for Use ...
Page 200: ...7 Specialty Transducers 7 6 Instructions for Use ...
Page 202: ...8 Physiologic Function 8 2 Instructions for Use ...
Page 208: ...9 eSieFusion Imaging 9 2 Instructions for Use ...
Page 236: ...10 Virtual Touch Applications 10 2 Instructions for Use ...
Page 258: ...10 Virtual Touch Applications 10 24 Instructions for Use ...
Page 302: ...Appendix A Technical Description A 44 Instructions for Use ...
Page 326: ...Appendix B Control Panel and Touch Screen B 24 Instructions for Use ...
Page 328: ...Appendix C Control Panel C 2 Instructions for Use ...
Page 394: ...Appendix D On screen Controls D 50 Instructions for Use ...
Page 444: ...Appendix F Acoustic Output Reference F 2 Instructions for Use ...
Page 516: ...Appendix F Acoustic Output Reference F 74 Instructions for Use ...
Page 517: ......
Page 518: ......