ACUSON
Freestyle User Manual
2-25
Safety and Care
Caring for the Ultrasound System
It is the responsibility of the user to verify that the ultrasound system is safe for diagnostic
operation on a daily basis. Each day, prior to using the system, perform each of the steps in the
Daily Checklist.
All exterior parts of the ultrasound system, including the control panel, keyboard, system
accessories, probes, and probe accessories, should be cleaned and/or disinfected as
necessary or between uses. Clean each component to remove any surface particles. Disinfect
components to kill vegetative organisms and viruses.
Daily Checklist
WARNING:
To minimize the risk of cross contamination and infectious diseases, a sterile,
non pyrogenic probe sheath must be in place during procedures requiring sterility.
WARNING:
To avoid electrical shock, you must visually inspect a probe prior to use. Do
not use a probe that has a cracked, punctured, or discolored casing or a frayed probe
adapter cable.
Discoloration Exception
: The use of the approved disinfectants, Cidex OPA and Gigasept FF,
may cause discoloration of probe housings, including the face of the probe. You can continue
to use a probe if it is discolored due to the use of these specific disinfectants only.
Perform the following each day before using the ultrasound system:
■
Visually inspect all probes. Do not use a probe that has a cracked, punctured, or
discolored casing or a frayed probe adapter cable.
■
Visually inspect all power cords. Do not power on the ultrasound system if a cord is frayed
or split, or shows signs of wear.
■
Verify that the Trackball, Rotary knobs, Soft Keys, and other controls on the control panel
appear clean and free from gel or other contaminants.
Once the ultrasound system is powered on:
■
Visually check the on screen displays and lighting.
■
Verify that the monitor displays the current date and time.
■ Verify that the probe identification and indicated frequency are correct for the active probe.
Summary of Contents for ACUSON Freestyle
Page 4: ...iv ACUSON Freestyle User Manual ...
Page 14: ...1 2 ACUSON Freestyle User Manual Introduction ...
Page 70: ...3 2 ACUSON Freestyle User Manual Setup ...
Page 78: ...3 10 ACUSON Freestyle User Manual ...
Page 80: ...4 2 ACUSON Freestyle User Manual Patient Studies ...
Page 92: ...5 2 ACUSON Freestyle User Manual Scanning ...
Page 110: ...6 2 ACUSON Freestyle User Manual Wireless Probes ...
Page 122: ...7 2 ACUSON Freestyle User Manual Batteries ...
Page 128: ...8 2 ACUSON Freestyle User Manual Connectivity ...