Scanning
Color Flow Mode Scanning
Pressing
Color
on the real-time Control Bar activates color flow mapping mode. This key
switches color flow on and off.
In color flow mode the real-time controls include:
C Gain
– Color Gain controls the color flow overall gain.
C Box
– Color Box controls the up/down position of the color flow box.
Color Flow Tools
The real-time Control Bar Tools Soft Key displays a number of secondary-level real-time
imaging-related controls. The active Tools items will vary according to the scanning mode.
Press the
Tools
button to display the Tools options. When this list is active, the Rotary knobs
will control interaction with the list. Use the knobs in the same manner as is used to control the
real-time control bar. The outer knob moves up and down the list and the inner knob adjusts
the value or state for individual items.
Color Flow Mode Tools Sub-Menu
Color Flow Mode
Tools Sub-Menu
Description
Exit Tools
Exits the
Tools
sub-menu.
Color Map
Steps through multiple color flow maps. This selection also allows you to move from
velocity mode to power mode.
They are displayed in the on-screen settings window as either
Vel
(velocity maps) or
Pwr
(Amplitude Power maps). The maps are represented in the color bar next to the real-time
image.
Vel Invert
Velocity Invert
. Shifts the color map orientation from the default of red = towards the
transducer, to blue = towards the transducer. When invert is selected, the color map
graphic bar will invert.
Displayed in the on-screen settings window as
Vel
.
WARNING:
It is important to pay careful attention to the color flow invert
status of the system. The non-inverted state is red=towards the transducer,
displayed at the top of the color bar and blue=away from the transducer,
displayed at the bottom of the color bar.
Priority
Adjusts the arbitration that takes place between B-Mode tissue signals and Color Flow
mapping signals. Higher priority settings will give greater weight to color flow signals. This
can be used with weaker flows or to fill in flow near vessel borders.
Displayed in the on-screen settings window as
Pri
.
C Persist
Color Persistence
. Increases or decreases the amount of persistence applied to Color
Flow signals. Higher settings will produce a smoother looking image and may help with
weak flow, but may have an appearance of lower frame rates.
Displayed in the on-screen settings window as
CP
.
C Filter
Color Filter
. Adjusts the Color Flow high pass filter. Lower settings will pass through
lower frequency flow signals.
Displayed in the on-screen settings window as
CF
.
C Scale
Color Scale
. Increases or decreases the Color Flow pulse repetition frequency. The
scale is displayed next to the color bar in centimeters per second.
Summary of Contents for ACUSON Freestyle
Page 4: ...iv ACUSON Freestyle User Manual ...
Page 14: ...1 2 ACUSON Freestyle User Manual Introduction ...
Page 70: ...3 2 ACUSON Freestyle User Manual Setup ...
Page 78: ...3 10 ACUSON Freestyle User Manual ...
Page 80: ...4 2 ACUSON Freestyle User Manual Patient Studies ...
Page 92: ...5 2 ACUSON Freestyle User Manual Scanning ...
Page 110: ...6 2 ACUSON Freestyle User Manual Wireless Probes ...
Page 122: ...7 2 ACUSON Freestyle User Manual Batteries ...
Page 128: ...8 2 ACUSON Freestyle User Manual Connectivity ...