ACUSON Cypress™ Operator’s Manual
1-11
External Devices and Connections
External Devices and Connections
The Cypress system has a number of external connections. For further details
see the Appendix “Connecting External Devices” on page A-1.
FIGURE 6.
External Connections
WARNING
Do not use a peripheral device, such as a monitor, VCR or printer within
1.5 meters of a patient unless the device receives power from an isola-
tion transformer that meets medical safety standards.
WARNING
Accessory equipment connected to the analog and digital interfaces
must be certified according to the respective IEC standards (IEC 950 for
data processing equipment and IEC 601-1 for medical equipment). Fur-
thermore, all configurations shall comply with the system standard IEC
601-1-1. Any party who connects additional equipment to the signal
input part or signal output part is considered to be configuring a medical
system, and is therefore responsible for ensuring that the system com-
plies with the requirements of the system standard IEC 601-1-1. If you
have additional questions, contact your Siemens representative.
1
See Appendix B for information about ink-jet printers.
15-PIN CONNECTOR FOR VCR,
VIDEO PRINTER, OR EXTERNAL
MONITOR
ONE UNIVERSAL SERIAL BUS
(USB) PORT FOR CONNECTING
AN EXTERNAL KEYBOARD,
EXTERNAL MOUSE, INK-JET
PRINTER
1
, OR USB HUB
RJ-45 JACK FOR 10/100 Mbps
ETHERNET CONNECTION AND
STATUS LEDs
PS2 PORT FOR CONNECTING
AN EXTERNAL KEYBOARD
MICROPHONE JACK
SECURITY LOCK