Parameter assignment
7.13 P053 Vessel dimension L
SITRANS Probe LR with mA/HART
Operating Instructions, 05/2019, A5E32337711-AC
49
7.13
P053 Vessel dimension L
Dimension L as used in P050 Vessel Shape.
Values
Range
0.0000 to 99999 in units (P005)
Default
0.0
Related parameters
P050 Vessel Shape
Enter the vessel length L (excluding both end sections) if P050 = 7. Use the units selected in
P005.
7.14
P054 Level breakpoints
When the vessel shape is too complex for any of the preconfigured shapes, you can define
the shape as a series of segments. In P054 you assign a level value to each breakpoint. In
P055 you assign a corresponding volume value to each breakpoint.
Primary index
P054
Secondary index
Breakpoint number
Values
Range
0.0000 to 99999 in units (P005)
Default
0.0
Related parameters
P055 Volume Breakpoints
Enter up to 32 level breakpoints, where the corresponding volume is known. The 100% and
0% levels must be entered. The breakpoints can be ordered from top to bottom, or the
reverse.
1.
First set P050 to 9.
2.
Select P054.
3.
The empty breakpoint value field appears, with the breakpoint number 01 in the auxiliary
reading.
①
Breakpoint value field
②
Breakpoint number
4.
Key in the level value for breakpoint 1, and press ENTER
. (Use units defined in
P005.)
5.
Press DISPLAY
twice to focus control on the secondary index.
6.
Press the UP ARROW
key to display 02 in the auxiliary reading.