Diagnostics and Troubleshooting
9
9.1
Fault codes
Error system
MAG 8000 can detect and report 13 different faults.
The faults are divided into two types: Fatal errors and Warnings.
Fatal errors: Faults 1, 2, 3, and 4
Warnings: Faults 5, 6, 7, 9, L, E, C, d, and 14
Fault
codes
Name/text
Description
Cause
Remedy
1
Insulation fault "Cross-talk" between
coil circuit and electro‐
des. An insulation fault
will result in a offset er‐
ror on flow measure‐
ment
(Advanced version on‐
ly)
● Defect cables between
sensor and transmitter
● Wrong connections of
remote cables
● Water between coils
and electrodes – could
be caused by a broken
liner
● Pollution/moisture on
PCB caused by water in
encapsulation
Check cable and wiring installation and dis‐
turbing elements in the surrounding.
Re-activation of insulation test (parameter
800) will reset alarm and start a new insu‐
lation test.
During the 4 minute insulation test, meas‐
urement is disabled.
Alarm remains active until a new test period
detects no failures.
2
Coil current
fault
Error in the coil circuit.
Coil current has not set‐
tled within specified pe‐
riod.
A short-circuit in the coil
can NOT be detected
● Coils have been discon‐
nected could be caused
by broken cable or wires
Check cable and wiring installation.
Alarm remains active until the fault condi‐
tion is corrected.
3
Amplifier over‐
load
Input signal is outside
expected range.
Input amplifier circuit
can not provide a stable
measurement.
A short circuit between
the two electrodes or
between a electrode
and common can NOT
be detected
● Electrodes have been
disconnected, or con‐
nected to ground
Check cable and wiring installation.
Alarm remains active until the fault condi‐
tion is corrected.
SITRANS FM MAG 8000
Operating Instructions, 03/2020, A5E03071515-AH
87
Summary of Contents for 7ME6810
Page 16: ...Description 3 4 Benefits SITRANS FM MAG 8000 16 Operating Instructions 03 2020 A5E03071515 AH ...
Page 140: ...Unit conversion table SITRANS FM MAG 8000 140 Operating Instructions 03 2020 A5E03071515 AH ...
Page 168: ...Features SITRANS FM MAG 8000 168 Operating Instructions 03 2020 A5E03071515 AH ...