Installing/Mounting
4
SITRANS F flowmeters with minimum IP65/NEMA 4X enclosure rating are suitable for indoor
and outdoor installations.
● Make sure that pressure and temperature specifications indicated on the device nameplate /
label will not be exceeded.
WARNING
Installation in hazardous location
Special requirements apply to the location and interconnection of sensor and transmitter.
See Installation in hazardous area (Page 15)
4.1
Installation safety precautions
WARNING
High pressure hazard
In applications with working pressures/media that can be dangerous to people, surroundings,
equipment or others in case of pipe fracture, we recommend that special precautions such as
special placement, shielding or installation of a pressure guard or a safety valve are taken
when the flowmeter is mounted.
Note
Install the sensor in well-supported pipelines in order to support the weight of the flowmeter.
SITRANS FM MAG 5100 W
Operating Instructions, 01/2020, A5E03063678-AC