Functions
7.5 Setting Relays
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
89
1.
FAULT – When selected the system will declare the Reflexor channel to be in Fault and
the flow meter will indicate "F" on the display screen. Note that when a new Reflexor
channel is created the [Fault] option is the default mode. Also note that the sum (or
difference) channel will also be declared to be in Fault.
2.
ZERO – When selected the Reflexor channel will be declared to be at Zero Flow instead
of in Fault and the flow meter will indicate 0.00 on the display screen.
7.5
Setting Relays
Relay Functions
Use the [Relay Setup] menu to assign a function to channel relays. The flow meter supports
two types of relay outputs, Alarm Relay and Pulse Relay. Alarm Relay outputs operate in
"fail-safe" mode. The relay(s) are energized under normal conditions - an alarm condition
causes the relay(s) to de-energize until the alarm clears. The Pulse Relay output supports
Totalizer and batch relay functions, with an output pulse width of approximately 200 ms;
maximum activation rate is 2.5 pulses per sec. If Totalizer pulses exceed this rate, excess
pulses are stored in an overflow register. This allows the relay to "catch up" when flow
decreases enough.
Note
Using the <F1> key (Totalizer clear command) also clears all channel Totalizers plus the
overflow register described in the last paragraph.
Relay 1 and 2 Function Assignments
The flow meter, depending upon the model, provides at least two alarm relays. Please refer
to the Appendix (Page 143) for wiring details. Relays respond to any of the alarm conditions
or data functions included on the Relay Option List.
Table 7- 4
Relay Option List
Not Used
Not Active
Power Off
Power Off alarm occurs when power fails.
High Temp
High temperature value relay trip-point.
Low Temp
Low temperature value relay trip-point.
High Flow
System flow rate exceeds high flow set point.
Low Flow
System flow rate falls below low flow set point.
Flow Alarm
System flow rate exceeds or falls below flow set points.
Fault Alarm
System loses receive signal (all paths in fault).
Soft Fault
Fault condition - memory mode active.
Spacing
System sensor spacing needs adjusting.
Empty
Empty pipe alarm.
Aeration
Aeration percentage exceeds alarm set point.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......