Functions
7.3 Logger Control
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
85
4.
To transmit Logger contents to external device via the serial port press <ENTER>.
5.
To stop printout press <Left Arrow>.
Circular Memory
In its default mode, the Logger collects data until its memory becomes full. At that time the
flow meter suspends datalogging and cannot resume until the Logger memory is cleared
(see Clear Logger command). Circular Memory allows the Logger to "‘write over" its oldest
records when memory reaches full capacity. If you enable [Circular Memory], you are
assured of always collecting the most recent data. But also remember that you will lose the
oldest Logger reports and that further invoking of [Circular Memory] deletes the current
contents of the Logger.
To setup and enable Circular Memory:
1.
The Logger Mode menu must have the [Memory] menu cell selected.
2.
Logger items must be selected (e.g., Site ID, Date, Time, etc.).
3.
All active channels/paths in the Channel Setup menu must be disabled. To disable active
channels, select the [Channel Enable] menu cell and then [No].
4.
In the Logger Control menu, select [Circular Memory].
5.
Press <Right Arrow> to access the [Circular Memory] option list.
6.
Move the cursor to [Yes] by pressing <Up/Down Arrow>.
7.
To store selection press <ENTER>.
8.
Lastly, re-enable the channels/paths that you disabled earlier to begin logging.
Est LogTime Left
Est LogTime Left is a "view-only" menu cell that shows an estimate of the amount of Logger
time remaining in hours and minutes. This menu cell becomes active after you enable
datalogging. Selecting [Circular Memory] and/or event-based datalogging (see Logger
Setup), blanks the [Est LogTime Left] field and is based on the log interval and data
selections made in the Logger Setup.
Clear Logger
If you use the Logger in its default mode, eventually you will use all the memory available for
Logger storage. When this occurs, you will not be able to log more data until you free up the
memory. The [Clear Logger] command erases ALL stored Logger data. Therefore, you
should evaluate the currently stored data, and print any valuable information before using
this command.
Note
Saved Sites also consume Logger RAM.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......