Functions
7.2 Span Data
FUP1010 IP67 Portable
Operating Instructions, 03/2012, A5E02951522-02
81
7.2
Span Data
The [Span Data] menu allows you to set 0% and 100% output limits for volumetric flow (Vfo),
absolute flow (Vfab) and sonic velocity (Vs). Each menu cell shows appropriate rate units
and time base. If you change flow rate units after spanning the system, the computer
automatically updates the output data setup to reflect the change. Span limits apply to both
the analog outputs and the on-screen strip chart. The flow outputs operate as follows:
Vfo
Spanned Volumetric::
The minimum and maximum flow rate entries
establish the Vfo span. The Max Flow menu cell
sets 100% of span. The Min Flow menu cell sets
0% of span. Use signed numbers for bi-
directional spanning. Note that negative (reverse)
flow always is lower than positive flow, whatever
its absolute magnitude. For example, for a flow
measurement range of -30 GPM to +10 GPM, the
4mA span will be -30 GPM, and the 20mA span
will be +10 GPM.
Vfab
Spanned Absolute Volumetric Flow Rate:
Vfab is the absolute magnitude of the volumetric
flow rate (Vfo). There are no menu cells provided
to span this output. Vfab shares the Vfo span
entries. The Vfab minimum span is always zero.
The maximum span for Vfab is the largest
absolute value of either the min. or the max. flow
rate (Vfo) entries. For example, a span between
+10 GPM and -30 GPM, spans the Vfab output
from 0 GPM to 30 GPM.
Vs
Spanned liquid sonic Velocity:
Vs is the sonic velocity in meters-per-second
(m/s) of the flowing liquid. The min. and max. Vs
entries establish the Vs span. Max Vs (m/s)
defines 100% of span. The Min Vs (m/s) defines
0% of span.
Maximum span values represent:
Minimum span values represent:
100% of span
Current output of 20mA
Voltage output of 10 VDC
Pulse output of 5000 Hz
0% of span
Current output of 4mA
Voltage output of 0 VDC
Pulse output of 0 Hz
To change the default Span Data settings:
1.
At [Meter Type], press <Right Arrow> to [Dual Path Flow] and press <ENTER>.
2.
Scroll to [Span/Set/Cal] and press <Right Arrow>.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......