Description
3.3 Theory of Operation
FUP1010 IP67 Portable
18
Operating Instructions, 03/2012, A5E02951522-02
a) ρ
60
= (141.5 *999.012) / (131.5 + API60)
●
where:ρ
60
= density at 60
o
F (Kg/m
3
)
b) α
60
= K
O
/ ρ
602
+ K
1
/ ρ
60
●
where K
0
and K
1
are coefficients of thermal expansion.
c) ρ
t
=ρ
60
* EXP (- α
60
Δt (1 + 0.8 α
60
Δt))
●
where Δt = (t - 60)
o
F and t = measured temperature
d) Therefore the volume correction factor (VCF) = ρ
t
/ ρ
60
The final volume compensated flow rate is then: Std Flow = FLOW
BULK
* VCF
Flow meter Types
The flow meter automatically conditions Installation Menu choices to suit the selected meter
type. The following illustrates the 2-Channel and 2-Path flow meter types.
2-Channel
Dual Channel provides two independent measurement channels that operate
simultaneously. Depending on the specific model, Dual Channel supports: Clamp-on Transit-
time, In-line Transit-time and Reflexor.
①
⑤
Upstream Sensors
Pipe B Output Flow
②
Downstream Sensors
⑥
Channel 2 for Pipe B
③
Channel 1 for Pipe A
⑦
Pipe B
④
Pipe A Output Flow
⑧
Pipe A
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......