Appendix
A.4 Flowrate Calibration and Calibration Tables
FUP1010 IP67 Portable
152
Operating Instructions, 03/2012, A5E02951522-02
3.
Press the <Right Arrow> and then the <Down Arrow> to switch the option list to: [Clr
Saved Data? Yes].
4.
Press <ENTER> to clear all Saved Site Data, Datalogger Data, User created Pipe Data
and Sensor Data.
5.
To restore operation, press <MENU> to access the Installation Menu and create a new
site setup. Reselect any Meter Facilities items (e.g., RS-232 setup parameters).
A.4
Flowrate Calibration and Calibration Tables
Flowrate Calibration Methods
SITRANS F 1010 equipment provides three ways to condition the calibration performance of
its flowrate output: Intrinsic (factory set), Kc, and Calibration Tables 1 through 3. Access to
these calibration options is found in the [Calibrate Flowrate] menu cell and the three [Calib.
Table] menu cells of the [Span/Set/Cal] menu.
Intrinsic
When selected, the flow meter uses no slope adjustment at all. Output data is still zeroed
and corrected for Reynolds number, but no slope adjustment is imposed on the flow meter’s
flow register.
Some applications may require an output adjustment to match an official external reference.
The [Calibrate Flowrate] menu allows you to select a calibration mode. The right-hand
column shows the active calibration mode. You can select Intrinsic (factory) and Kc (Slope
Correction) Calibration. Selecting either of the external calibration modes will not eliminate
the Intrinsic (factory) calibration. You can use this menu cell to switch between Intrinsic and
Kc at any time.
Kc Calibration
For most applications, the measured flow range produces a linear meter response.
Therefore, the Kc (slope correction) calibration is the preferred method since it only requires
a single correction factor for all the flow rates encountered.
Note
Changing the calibration can cause profound changes in flow meter operating
characteristics. Use only the most respected flow standard to obtain a correction factor. The
percentage entered must provide an accurate and consistent shift across the entire flow
range anticipated for the application.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......