Troubleshooting
10.3 Test Facilities Graph Screen
FUP1010 IP67 Portable
134
Operating Instructions, 03/2012, A5E02951522-02
Signal Masking Function: (Hot Key 7)
Under conditions of extremely low signal amplitude, a noise spike associated with the flow
meter receive signal window may be present on the extreme left side of the graph display. If
this spike is large enough it may interfere with the signal detection routines.
1.
To eliminate this noise from the signal processing routines, press the <7> key to invoke
the [Mask Set #] command, then press the <+> key until the noise is no longer present in
the receive waveform.
2.
Press <0> to exit this command.
Hold Set Function: (Hot Key 8)
The [Hold Set #] command is used to set the Hold Set number higher if intermittent mis-
registration occurs. Press the <8> key on the keypad to invoke this function.
Table 10- 2 Description of Graph Screen Text Display Parameters
Screen Text Parameters
Menu List Item
Description
F
Flow
Measured flow rate in selected flow
units.
VS
Vs m/s
Sound Velocity in meters per second.
[ ]
Display Metrics
Represents the digital sample position
of the receive window.
Correlated plot
Displays the receive waveform in its
proper superposition or registration.
The true delta time will be displayed by
NOT selecting "Correlated Plot".
Centroid Mark
Indicates with a large vertical marker
the peak energy of the receive
waveform.
D
Damping
Displays the minimum and maximum
digital damping exponent along with the
active damping exponent.
Tn
Tn (usec)
Receive signal transit time in
microseconds.
dT
DeltaT (nsecs)
Transit time delta (difference) in
nanoseconds.
S
Signal Strength
Displays %Valc (amplitude), %Vaer
(aeration factor) and numeric ALC.
Signal-to-Noise Ratio
Indicates the signal to noise ratio of the
receive signal. Increased damping will
increase the S/N ratio as the
asynchronous noise reduces.
SN
Envelope
Percentage change of the signal from
Initial Makeup conditions.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......