Troubleshooting
10.3 Test Facilities Graph Screen
FUP1010 IP67 Portable
126
Operating Instructions, 03/2012, A5E02951522-02
Clearing All Saved Data
1.
Turn off power (if it is currently on).
2.
Press <F4> and keep it pressed while you turn on power. The prompt: [Clr Active
Memory? No] appears at the top of the screen. Press the <Down Arrow>. Note that the
prompt switches to [Clr Saved Data? No].
3.
To access the F4 Reset option list press the <Right Arrow>. Press the <Down Arrow> to
switch the option list to [Clr Saved Data? Yes].
CAUTION
Before proceeding further it is essential to understand that this function eliminates ALL
data stored in RAM. This means that all saved site setups including the site data of a
flow-calibrated site will be erased! In addition, the entire Datalogger file plus any custom
factory or user-created pipe or sensor tables will be eliminated. The impact of this is
such that we strongly recommend that you consult Technical Services before continuing
with this procedure. Be aware that you will have to create a new Site Setup, re-enter all
site specific parameters including pipe or sensor tables, plus all desired Meter Facilities
menu entries.
4.
To clear all Saved Memory press <ENTER>.
5.
Create a Site Setup before attempting to access other menu items.
6.
To restore operation, press <MENU> to access the installation menu. Create a new site
setup and complete the installation procedure.
7.
Re-select desired Meter Facilities menu items (e.g. RS-232 setup parameters).
10.3
Test Facilities Graph Screen
When operating in the transit time mode the Test Facilities Graph Screen is an exceptional
diagnostic tool for troubleshooting problem applications or simply determining Receive signal
quality. The primary function of this screen is to display the digitized receive signal waveform
with the similar appearance and function of a digital oscilloscope. This screen also allows the
user to override some of the flow meter default settings by permitting adjustment to the
measured transit time, the digital averaging and the zero crossover used in the
measurement of the up/down transit time difference. The figure shown below is a
representation of the diagnostic graph.
Note
The Test Facilities Graph Screen requires significant CPU overhead. The flow meter should
not be left in this mode during normal operation where the Datalogger is the primary output
or during calibration work.
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......