Functions
7.10 RTD Calibration
FUP1010 IP67 Portable
100
Operating Instructions, 03/2012, A5E02951522-02
5.
Press <ENTER>to register setting. This adjusts the meter’s DAC (digital-to-analog
converter) so that a 1 kHz output corresponds with 1 kHz on the frequency counter.
6.
Recheck the frequency counter to make sure that it is now reading 1 kHz.
7.10
RTD Calibration
Resistive Temperature Device (RTD) Calibration
The RTD Calibrate Menu appears on all SITRANS F 1010 models. Use this menu to
calibrate Temperature Sensors to an external standard. It is important to note that Siemens
RTD temperature sensors are factory-calibrated for high accuracy. We recommend that
before deciding to perform the calibration, check the current RTD reading in the
[Diagnostics/Liquid Data] menu. You may find that you do not even need to calibrate the
sensor. In any case, make sure that the temperature reading stabilizes before proceeding
further. The RTD Calibrate menu allows you to perform an external calibration, which can be
accomplished either by data entry of the current RTD temperature or by a 0°C (32°F) Ice-
Bath procedure. You can switch between the intrinsic and external calibration modes at any
time.
Note
If you perform an external temperature calibration, you should mark and record the location
of each connector and sensor-cable. Once you have re-calibrated the temperature sensors,
changing the sensor/connector orientation established during the procedure may void the
calibration.
1.
From the [Meter Facilities] menu scroll to the [RTD Calibrate] menu.
2.
To access the [RTD Calibrate] menu press <Right Arrow>.
0HWHU)DFLOLWLHV
6LHPHQV
'XDO3DWK>@3DWK
&DOLEUDWH7HPSHUDWXUH6HQVRUV
3UHIHUUHG8QLWV
(QJOLVK
7DEOH6HWXSV
/RJJHU&RQWURO
0HPRU\&RQWURO
$QDORJ2XW7ULP
57'&DOLEUDWH
&ORFN6HW
566HWXS
1RQH>@
%DFNOLJKW
2Q
6\VWHP,QIR
Summary of Contents for 7ME3510
Page 2: ......
Page 10: ...Table of contents FUP1010 IP67 Portable 8 Operating Instructions 03 2012 A5E02951522 02 ...
Page 159: ...1010WP 7 21614 C ...
Page 160: ...1010WP 7 21614 C ...
Page 161: ...1010WDP 7 21614 C ...
Page 162: ...1010WDP 7 21614 C ...
Page 163: ......
Page 164: ......
Page 165: ......
Page 167: ...1012FP 8 DIRECT MODE REFLECT MODE INSTALLED ON PIPE 21614 C ...
Page 168: ......
Page 169: ...21614 C ...
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ...21614 C ...
Page 175: ...21614 C ...
Page 176: ...21614 C ...
Page 177: ...21614 C ...
Page 178: ...21614 C ...
Page 179: ...21614 C ...
Page 180: ...21614 C ...
Page 181: ......
Page 182: ......
Page 183: ...21614 C ...
Page 184: ...21614 C ...
Page 185: ...1012TP S 8 OUTLINE DIMENSIONS REFLECT MODE DIRECT MODE 1012TP S SERIES 21614 C MOUNTING TRACK ...
Page 186: ...REFLECT MODE DIRECT MODE 21614 C ...
Page 187: ...1015BC 1 8 21614 C ...
Page 188: ...21614 C INPUT OUTPUT TERMINALS 1015WP T10 8 ...
Page 189: ...21614 C INPUT OUTPUT TERMINALS 1015WP T26 8 ...
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 195: ...DUAL HEAD CONFIGURATION 21614 C IN LINE CONFIGURATION FLOW ...
Page 196: ......
Page 197: ......
Page 208: ...Glossary FUP1010 IP67 Portable 162 Operating Instructions 03 2012 A5E02951522 02 ...
Page 212: ...Index FUP1010 IP67 Portable 166 Operating Instructions 03 2012 A5E02951522 02 ...
Page 213: ......