Operation
5.4 Operating functions
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
87
5.4.6.5
ELAN configuration (function 73)
73 ELAN Config.
O
2
Channel address:
O1
Measured value
telegram : off
ELAN-TAG : OXY64
Figure 5-51 Configuring ELAN
With this function, you set the parameters for an ELAN network.
Channel address:
Set the channel address here. Addresses from 1 to 12 can be set. In an ELAN network, each
address may only be used once.
Measured value telegrams on/off:
Here you can switch the cyclic, independent sending of measured values every 500 ms on or
off.
ELAN tag:
Display of the tag assigned to the analyzer in the ELAN network. It has a maximum of 10
alphanumeric characters.
Note
Independent sending of measured values
When you have set up your own communication control system, this option offers a simple
way of checking an ELAN telegram. In order to avoid unnecessarily loading the device and
the ELAN network, however, the function should only be switched on as needed!
Further ELAN details can be found in the ELAN interface description:
● (C79000-B5200-C176 German)
● (C79000-B5276-C176 English).