
Operation
5.4 Operating functions
OXYMAT 64
58
Operating Instructions, 06/2008, A5E00880383-03
5.4.3.6
Autocal (function 24)
Note
You can only make use of automatic calibration (Autocal) if your device contains additional
electronics (optional board). If this isn't the case, a corresponding message will appear on
the display when the Autocal function is called.
Automatic calibration can only be started if the device is in the "Measure" operating state!
Autocal/-check O
2
Figure 5-14 Autocal function
Note
The settings for "Autocal/Check..." (subfunctions 1 to 3) are valid both for Autocal (automatic
calibration) and for Autocal Check (automatic check that the set calibration tolerances are
complied with, without calibration).
The settings for "Autocal Check" (4th subfunction) only refer to the check of the calibration
tolerances without calibration.
AUTOCAL/Check operating mode
Autocal/-check mode O
2
Figure 5-15 AUTOCAL/Check operating mode