SPECIFICATIONS
10.4.2
Measured value recording
10.4.3
Ambient conditions
Measuring ranges (scalable)
– In-situ and extractive version
– Version for ambient gas measurement
0 … 21 vol% O
2
2 … 21 vol% O
2
Precision
±0.2 vol% O
2
Temperature dependence in T-range
±2 vol% of measured value, max. dT/dt 1 °C/min
Stability
Zero point drift ±0.1 vol% O
2
/ year
Reference gas drift ±0.8 vol% from measured value /
year
Measurement response time (T
63
/T
90
) in
still
air
– without filter
– with stainless steel mesh
– with stainless steel mesh and PTFE
10 s / 20 s
10 s / 25 s
30 s / 70 s
Operating pressure range
0.8 … 1.4 bar(a)
Start time
2.5 min
Warming up time (according to specification)
3 min
Display
7-segment LCD
LED
Three-colored: Red/yellow/green
Operation location
Outdoors or indoors.
No direct sunlight.
Use a weather hood when necessary.
Operating temperature range
– for probe (installed in process)
– for electronics (housing)
– for TRANSIC151LP voltage supply unit
– for TRANSIC151LP (ambient air measurement)
−
20 … +80 °C
−
20 … +60 °C
−
20 … +60 °C
−
20 … +60 °C
Storage temperature range
−
40 … +80 °C
Operating pressure range
0.8 … 1.4 bar(a)
Air humidity
100% r.h. non-condensing
Altitude
Up to 2000 above sea level
Electrical compliance
In accordance with DIN EN IEC 61010-1:2011
Safety information
Laser product of protection class 1; see
operating information”, page 8