MAINTENANCE
10.2
Operating state - Maintenance
▸
Either activate the Maintenance mode via the key switch “Maintenance” on the outside
of the system cabinet door. The sample gas pump remains active and a maintenance
signal is set.
or
▸
Activate Maintenance mode on the BCU. LED “MAINTENANCE REQUEST” lights. Menu
function “Maintenance signal”
activates status “C”.
When this status controls a digital output of the GMS800, an external source receives a
signal that the GMS800 is not in Measuring mode (e.g. during maintenance work). The
sample gas pump is switched off and the sample gas signals are not processed.
10.3
Visual inspection / functional check
▸
Perform outer visual check daily/weekly.
Pay attention to the signal lamps on the system cabinet door.
▸
Check the following components of the complete analysis system regularly (monthly):
– Screw connections
– Hose connections
– Gas lines
– Connections of the PowerCEMS100
– Connections of the probe
WARNING
TOXIC CONDENSATE
Serious damage to environment, nature and health.
▸
The liquid contained in the condensate container is acidic, alkaline
and contains inorganic or organic substances toxic or harmful to the
environment.
This waste must be disposed off according to legal regulations and
as hazardous waste when necessary.
▸
Wear personal protective equipment (gloves, respiratory protection).
Refer to the relevant Operating Instructions of the relevant system components
for details on maintenance and safety information.