Operating Instructions
Chapter
8
CLX490
8009237/RB47/2007-08-03
©
SICK AG · Division Auto Ident · Germany · All rights reserved
8-5
Troubleshooting
8.3.3
Messages for errors accessing the external parameter memory
The CLX outputs further error messages, in addition to the general error messages in
Chapter 8.3.1 CLX without external parameter memory, Page 8-2
,
if problems arise when
the external parameter memory is accessed.
Message
Meaning
Possible cause
Remedy
1. "error: access to external
parameters not successful
internal parameters
loaded"
(Error type: 104, 105,
106, 111, 112,115)
•
Start option: S
TART
WITH
E
X
-
TERNAL
P
ARAMETERS
.
The CLX has
not
loaded the
external parameter set. In-
stead, it has loaded the inter-
nal parameter set to its RAM.
The CLX, however, does not
output data on the host in-
terface.
•
Start option: S
TART
WITH
C
OPY
I
NTERN
–> E
XTERN
.
The CLX was
not
able to copy
the internal parameter me-
mory. It has loaded the inter-
nal parameter set to its RAM.
The external parameter memory is
defective.
Repeat the startup procedure. If
the error is still present, contact
the SICK Service department.
2. "error: external parameter
memory full
internal parameters
loaded"
(Error type: 108)
Start option: S
TART
WITH
C
OPY
I
NTERN
–> E
XTERN
.
The CLX was
not
able to copy
the internal parameter memory.
It has loaded the internal para-
meter set to its RAM.
The internal parameter set is too
large for the external memory
(memory is full).
Contact the SICK Service depart-
ment.
3. "error: external parameters
not valid for device
internal parameters
loaded"
(Error type: 116)
Start option: S
TART
WITH
E
XTER
-
NAL
P
ARAMETERS
.
The CLX has
not
loaded the
external parameter set. Instead,
it has loaded the internal para-
meter set to its RAM.
The CLX,
however, does not output
data on the host interface.
The parameter set in the external
memory is not from a CLX490.
Use the correct external parame-
ter memory or overwrite the cur-
rent one.
4. "error: no external para-
meter memory connected
internal parameters
loaded"
(Error type: 102)
•
Start option: S
TART
WITH
E
X
-
TERNAL
P
ARAMETERS
.
The CLX has
not
loaded the
external parameter set. In-
stead, it has loaded the inter-
nal parameter set to its RAM.
•
Start option: S
TART
WITH
C
OPY
I
NTERN
–> E
XTERN
.
The CLX was
not
able to copy
the internal parameter set to
the external memory. It has
loaded the internal parame-
ter set to its RAM.
•
External parameter memory not
connected.
•
External parameter memory not
connected.
•
OK if the device is to be opera-
ted without an external para-
meter memory.
•
Check whether the memory is
connected correctly (free
wiring: check wire assignment).
•
See
5 Remedy 2nd paragraph
5. "error: writing parameters
CLV->EXT internal para-
meters loaded"
(Error type: 110, 111,
112)
Start option: S
TART
WITH
C
OPY
I
NTERN
–> E
XTERN
.
The CLX was
not
able to copy
the internal parameter set to
the external memory. It has
loaded the internal parameter
set to its RAM
Data transmission error.
Check whether the external para-
meter memory is connected cor-
rectly (free wiring: check wire
assignment).
Repeat the startup procedure. If
the error is still present, contact
the SICK Service department.
Table 8-3: For messages for problems accessing the external parameter memory