8.1
If the device does not function at all …
8.2
When measured values are obviously incorrect …
8.3
Malfunction messages
A malfunction has occurred when:
●
The measured values blink.
●
The “
yellow
” LED lights.
●
The “
red
” LED lights.
Press the <
Diag
> button for more information.
8.4
Stand-by
When a malfunction occurs, the device switches to “Stand-by“:
●
The sample gas path (including gas sampling system) is purged with instrument air.
Press the <
Diag
> button for more information.
Possible cause
Notes
Power supply has failed.
Check the power supply (e.g. external switch, external fuses).
Internal fuse defective.
Check fuses (
Software not working correctly.
Switch the device off with the external mains switch and switch
it on again after a few seconds.
Possible cause
Notes
The device does not measure the
sample gas.
Check the measured medium path and all valves (e.g. switch-
ing from the test medium to the measured medium).
Sample gas path not gas-tight.
Check the installations.
The device is not adjusted correctly.
Perform an adjustment (
p. 41, §5.4); check the test media
first (setpoint value, service life, throughflow, setting in menu
2.3.1).
More information concerning the causes:
Controls and indicators/displays
Classifications (device state)
List of error messages and possible causes
More information concerning the causes:
Classifications (device state)
List of error messages and possible causes