85
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS
5
Fig. 122: Contact plug, assignment with circuit diagram
13
Fasten the fastening clip in the contact plug.
14
Close the contact plug.
15
Fasten the cell covers (
see Chapter “Fitting the complete cell”, page 68
16
Fit the cell (
see Chapter “Fitting the complete cell”, page 18
).
5.2.7
Replacing the thermostatic switch
The thermostatic switch can be replaced when the cell is installed.
Fig. 123: Thermostatic switch
Tools required
Phillips screwdriver
Pliers
Slot screwdriver
Spare parts set No.
Contents
6053091
Thermostatic switch 235 °C (for cell)
DANGER: Risk of burns due to hot surface.
Allow the MCS200HW to cool down before carrying out any work.
1
Thermostatic switch
2
Release button