113
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS
5
11
Insert the filter wheel into the filter wheel housing and screw tight:
1
Insert with the open side facing downwards.
2
Align the filter wheel so that the fastening screw can be accessed through the
opening in the housing.
12
Insert the filter wheel housing into the sender/receiver unit and screw tight.
13
Connect the lug connection of the filter wheel housing to the LPMS04 circuit board.
14
Replace the drying agent (
see Chapter “Replacing the drying agent”, page 44
).
15
Close the sender/receiver unit cover and screw tight.
5.3.10.3
Filter wheel motor 3
1
Open the sender/receiver unit (
see Chapter “Opening the sender/receiver unit”,
).
2
Disconnect the plug connection on the LPMS04 circuit board.
3
Loosen the fastening screws of the filter wheel housing with a 5 mm Allen key (2 pieces).
Fig. 172: Filter wheel 3
The axle is chamfered.
The spare parts set contains the connecting cable with fork light barrier. If necessary,
the connecting cable can also be replaced.
Spare parts set No.
Contents
2099795
Filter wheel 3, stepper motor kit
Tools required
Phillips screwdriver
The drying agent must be replaced each time the sender/receiver unit is opened.
Pay attention to dust protection: Replace the electronics housing cover during work on
the sender/receiver unit.
1 Filter wheel housing 3
2 Fastening screw, right
3 Filter wheel 3, plug connection
4 Plug connection, LPMS03 detector circuit board