REPAIRS
5
Fig. 164: Filter wheel housing 1, rear side
6
Remove the filter wheel
:
1
Align the filter wheel in the housing so that the fastening screw can be accessed
through the opening.
2
Loosen the fastening screw with a 2 mm Allen key (loosen the screw only, do not
remove).
3
Remove filter wheel 1.
Fig. 165: Filter wheel 1, screw connection
1 Photoelectric switch
2 Motor and light barrier cable connection
3 Lens
4 Filter wheel motor
NOTE: Material damage
Do not touch the optical surfaces of the filter wheels.
Alternative:
a) Loosen two screws inside the filter wheel housing (
wheel housing 1, inner side”, page 108
b) Remove the filter wheel from the clamping hub (the clamping hub
remains on the motor shaft).
c) To remove the clamping hub from the filter wheel housing, loosen the
fastening screw (
see Fig. 166 “Filter wheel 1”, page 110
).
d) After removing the hub from the filter wheel housing, fasten it back on the
filter wheel.
e) Continue with step 7.