105
8024638/AE00/V1-0/2019-09| SICK
S E R V I C E M A N U A L | MCS200HW
Subject to change without notice
REPAIRS
5
5.3.9
Replacing the chopper motor
Fig. 156: Chopper motor
1
Disconnect the MCS200HW power supply.
2
Open the sender/receiver unit (
see Chapter “Opening the sender/receiver unit”,
)
3
Remove the complete beamer unit (
see Chapter “Replacing the complete beamer unit”,
, steps 2 to 6).
4
Loosen the fastening screw of the chopper wheel cover on the upper side (2 mm Allen
key).
Fig. 157: Chopper wheel
5
Remove the chopper wheel cover.
6
Loosen the fastening screw of the chopper wheel cover on the lower side (2 mm Allen
key).
Tools required
2 mm Allen key
5 mm Allen key
Spare parts set No.
Contents
2099792
Chopper motor kit
The drying agent must be replaced each time the sender/receiver unit is opened.
Pay attention to dust protection: Refit the sender/receiver unit cover during work on the
unit.
1 Chopper wheel cover
2 Fastening screw