Steps
1.
Fit the measuring probe with pre-fitted purge air attachment on the flange with
tube.
2.
Tighten the nuts.
NOTE
The installation steps of the measuring probe on the flange with tube are identical for
all measuring probes.
5.6.7
Putting the purge air unit into operation
NOTE
Electrical connections for the purge air unit, see Purge air unit Operating Instructions.
Preparation
Switch the power supply for the purge air unit on for a short time to check the function
and to remove any dust that may have penetrated the purge air hose.
Figure 9: Purge air connection on purge air attachment of measuring probe
1
GMP measuring probe: Purge air hose connection
Steps
1.
Connect the purge air hose of the purge air unit with a hose clamp to the purge air
bracket connection of the purge air attachment.
2.
Switch on the purge air supply.
✓
Purge air supply is now activated and protects the measuring system against
contamination and overheating.
NOTICE
Attach clearly visible warning information against unintentional switching off on all
switching devices with which the purge air units can be switched off.
MOUNTING
5
8009389/147I/V4-0/2021-03 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM35
25