Flow Calibration
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
79
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
3.3.6
Identification of key data in the meter configuration
To ensure the validity of the flow meter configuration, the following data of the Parameters
Report in the Manufacturers Data Report (MDR) has to be consistent with the information
on the "Meter information" page (
) in MEPAFLOW600 CBM:
●
Serial number of the meter,
●
Firmware version,
●
Meter factor,
●
Firmware CRC.
Open the "Meter information" page (select Meter / Information).
Figure 27
Identification information on the "Meter information" page in MEPAFLOW600 CBM
3.3.7
Setting the current gas condition
Before the calibration can be started, the process parameters "TemperatureFix" (Reg.
#7040) and "PressureFix" (Reg. #7041) must be set to the average conditions during the
calibration. These parameters can be edited in the "Meter calibration" wizard (
).
Open the "Meter calibration" wizard in MEPAFLOW600 CBM (select Tools / Meter
calibration).
Enter p and T in the corresponding fields.