9.3.5
Resetting the error volume counters
To reset the error volume counters, password level "authorized operator" is required.
9.3.6
Acknowledging a battery replacement
After a battery replacement the following message flashes on the LCD-display:
Since date and time of the FLOWSIC600 are invalid after switching the meter back on, they
have to be set to valid values. We recommend to use the clock synchronization function of
MEPAFLOW600 CBM (click the "Attention" Icon in the "Identification" box on the
"Information" page).
After setting date and time, the LCD message can be acknowledged. The acknowledgment
sets the life span cycle of the new battery to zero. Complete the following step at the LCD-
display:
+Ef 70 m³
-Ef 0 m³
Select the error volume display. Use the ENTER function to enter the
dialog for resetting the error volume counters.
<ENTER>
Counter reset
OK Cancel
Confirm the reset with the ENTER function. The error volume counter will
be reset to zero. The reset will be recorded in the logbook. Press C/CE to
cancel the reset.
<ENTER>
+EV 0 m³
-EV 0 m³
Description of a battery replacement, see Service Manual
INFORMATION 1030
LifeSpan Battery
Press DATA.
INFORMATION 1030
LifeSpan Battery
Apply the ENTER function two
times
Device state
E:x W:x I:x
Apply the ENTER function
Battery replaced
OK Cancel
(This message becomes visible for just one
second.)
WARNING:
Explosion Hazard
Do not replace the battery unless power has been removed or the area is
known to be non-hazardous.
Substitution of components may impair intrinsic safety.
Replacement battery must be of type Panasonic BR 2032.