Commissioning and operation
FLOWSIC30 · Operating Instructions · 8020999/173B/V1-4/2020-02 · © SICK Engineering GmbH
49
Su
bj
ec
t t
o
ch
an
ge
w
ith
ou
t n
ot
ice
4.2.7
Analog out
For a setup of the analog out, the primary value relations to 4 mA (LRV) and 20 mA (URV)
have to be assigned for the defined Primary Variable (PV).
The default PV is the volumetric flow s. c.
To assign the value for 4 mA, proceed as follows:
1
“1 Device setup” > “3 Setup” > “4 HART/Loop current” > “6 PV LRV” > select “2 PV
LRV”
2
Insert the assigned value for 4 mA (default is 0.000 Cum/h)
3
Press ENTER
4
Press SEND to write the value to the meter
To assign the value for 20 mA, proceed as follows:
1
“1 Device setup” > “3 Setup” > “4 HART/Loop current” > “6 PV URV” > select “1 PV
URV”
2
Insert the assigned value for 20 mA (default is 1600.000 Cum/h)
3
Press ENTER
4
Press SEND to write the value to the meter
The analog damping of the defined PV may be changed as follows:
1
“1 Device setup” > “3 Setup” > “4 HART/Loop current” > “8 PV Analog damp”
2
Insert the damping value (default is 1 s)
3
Press ENTER
4
Press SEND to write the value to the meter