Assembly and Installation
Su
bj
ect
to
ch
an
ge
with
ou
t n
otic
e
3.8.6.1
Setting the bus address at sender/receiver unit (bus systems only)
On bus systems (e.g. 2x1-path connected to one MCUP), the required bus address of a
sender/receiver unit (master only) have to be assigned by the software. In delivery
condition the bus address for every sender/receiver unit is set by hardware to address "0".
Hardware addressing is read in with the start of the SOPAS ET program and cannot be
changed at site.
3.8.6.2
Software addressing at sender/receiver units
As standard, the address have to be assigned using the SOPAS ET program. The prereq-
uisite for this addressing is that hardware address 0 is set (delivery condition). To change it,
do the following:
Connect the measuring system to a PC/laptop, start the SOPAS ET program and make-
contact with the device (see section 4.1.3).
Select the sensor type (e.g. "FL100 EX-S") in the "Scan result" tab and move to the "Proj-
ect" window.
In the "Login" menu, select the "Service" user level and enter password "service"
Set the measuring system to the "Maintenance" mode
Select the menu "Configuration/Device Parameters" and choose the field "serial inter-
face", set the bus address.
3.8.7
Fitting and connecting the Interface and I/O module options
3.8.7.1
MCUP in compact or medium size housing
Plug interface modules and module carriers for I/O modules onto the hat rail in the MCUP
(
and connect to the associated connection on the processor board with
the cable with plug-in connector (
). Then plug the I/O module on the module
carrier.
Connect I/O modules using the terminals on the module carrier (
the Profibus module using the terminals on the module and connect the Ethernet module
via a customer provided network cable.
●
Terminal assignment AO module
NOTICE:
●
After a possibly required change of addressing, the respective sender/
receiver units must be started a new (disconnect and reconnect supply volt-
age). The output assignments in the MCUP then have to be reconfigured
(see section 4.2).
●
The sender/receiver units must have different addresses. Identical
addresses of several units cause the aborting of the communication with
the MCUP!
WARNING:
No other parameters or settings may be changed while logged into access level
2 - "Service", except for those described in this section. All other changes to
settings may only be carried out using the access levels 0 - "Operator" or 1 -
"Authorized client". The manufacturer is not liable for any disruption caused by
incorrect settings made by the user in SOPAS.
For detailed information on the particular module options, see „Interface Docu-
mentation FLOWSIC100 Flare“.