94
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Main Menu Functions
10.5.4
Operational displays in the diagnosis function
The following displays appear in succession:
Alarm status (threshold value messages)
Temperature of the electronic section of the analyzer unit
D i a g n o s i s
T h r e s h o l d v a l u e
T h r e s H 2 N O
T h r e s H 1 Y E S
T h r e s L 1 N O
T h r e s L 2 N O
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Alarm status
1
2
Current measured value
Current measuring range (measuring range end value)
1
JA = activated / NEIN = not activated
2
If YES is displayed although the current measured value is inside the threshold value: The latching function is
activated, manual deactivation is necessary (
D i a g n o s i s
I n s i d e T e m p . E l e A n a l v
a c t u a l 3 2 . 8 C E L
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Current temperature (CEL = °C)
Current measured value
Current measuring range (measuring range end value)
If the current temperature exceeds 70 °C, a malfunction message is displayed
and the heating of the sensor block is switched off.
< C o n t i n u e >
< C o n t i n u e >
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...