Calibration
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
153
12.6
Checking calibration
12.6.1
Zero check
Function
This function is used to have the zero gas measured instead of sample gas for test
purposes. As long as the function is activated, zero gas supply is activated (internal
solenoid valve opens) but the EuroFID remains in measurement mode, therefore displays
the measured value of the zero gas.
Procedure
Open the menu: Main menu
Maintenance
System Parameters
Zero Check
12.6.2
Sensitivity test (calibration gas test)
Function
In the same way as during the “Zero check”, this function can be used to measure the span
gas instead of the sample gas. When the function is activated, the span gas supply is
automatically activated (internal solenoid valve opens), but the EuroFID remains in
measurement mode.
This condition terminates automatically when a programmed time has elapsed (approx. 5
minutes) or if the menu function is previously cancelled.
Procedure
Open the menu: Main menu
Maintenance
System Parameters
Calib gas check
The calculated “deviations” are automatically compared with the drift limit
values (can be set in the “Expert level” sub-menu). If the drift is greater, a
malfunction message is generated.
Standard settings:
– Drift > 20 %: Malfunction message “Service requir.”
– Drift > 50 %: Malfunction message “System failure”
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...