60
8012426/YWL2/3-0/2016-08| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100
Subject to change without notice
4
START-UP AND PARAMETER SETTINGS
4.2.3
Assigning the sender/receiver unit to the measuring location (in SOPAS ET)
The sender/receiver unit can be assigned explicitly to the respective measuring location.
The following steps are then necessary:
▸
Start the SOPAS ET program and connect to the measuring system (
the device via USB line”, page 55
).
▸
Select device file “DH SP100” and move it to the “Project Tree” window.
▸
Enter password level 1.
▸
Set the sender/receiver unit to “Maintenance”: Click “Maintenance sensor”.
Fig. 34: SOPAS ET menu: DH SP100/Maintenance/Maintenance
▸
Select the “Configuration / Application Parameters” directory and enter the desired data
in the “Mounting location” entry field under “Device Identification”.
Fig. 35:
SO
PAS ET menu: DH SP100/Configuration/Application parameter
The respective device type connected is displayed automatically