O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-3K
Subject to change without notice
MENUS
11
11.2
Configuring on the control unit display
Some configuration options can also be set directly on the control unit display. Some
important functions are explained here in more detail as examples. The numbers behind
the submenus refer to the numbering of the menus in the previous subsections.
11.2.1
Configuring analog outputs and inputs of the control unit
1
Set the control unit to “Maintenance” (1.1) and activate submenu “I/O Parameter” (1.4).
2
Select the setting of the “Analog output parameters” (1.4.1) or the “Analog input
parameters” (1.4.2) and enter the password (
) using the
control fields.
3
Set the desired value using the operating fields. Press “Save” to save in the device.
11.2.2
Assigning the control unit to the sender/receiver unit
1
Set the control unit to “Maintenance” (1.1) and activate submenu “I/O Parameter” (1.4).
2
Select the setting of the “MCU variant” (1.4.3) and choose the type “DUSTHUNTER S”.
3
Enter the password (
) using the operating fields and save the
selected type with “Saved”.
11.2.3
Entering the regression coefficients
1
Set the sender/receiver unit to “Maintenance” (2.1) and activate submenu “I/O
Parameter” (2.2).
2
Select the desired parameter and enter the password (
the operating fields.
3
Enter the determined coefficient (
see “Calibration for dust concentration measure-
) with the operating fields and save to the device with “Save”.