67
8024493/AE00/V1-0/2020-09| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-2K
Subject to change without notice
CONFIGURING
9
9.3.2
Connect to the MCUDH Ex-3K control unit via Ethernet (option)
Recommended procedure:
1
Configure the Ethernet module via manufacturer software and note the IP address.
2
Connect the laptop/PC with the Ethernet module.
3
Connect the switched-off control unit to the network cable (LAN).
4
Switch control unit on.
5
Start SOPAS ET.
6
“Search settings”
7
“Search by interface”
8
Make the settings:
– Ethernet communication (always clicked)
– USB communication (always clicked)
– Serial communication: Click
9
Enter the noted IP address.
10
A list of COM ports appears.
Specify the COM port of the DUSTHUNTER.
If you do not know the COM port:
see “Finding the DUSTHUNTER COM port”, page 81
11
Assign a name for this search.
12
Click “Finish”.
To connect to the measuring system via Ethernet, the integrated interface module
Modbus ASCII/RTU must be installed in the MCUDH Ex-3K control unit (
module (option) of the MCUDH Ex-3K”, page 59
). An RS485 line can be used to connect
a remote Modbus TCP/IP or Ethernet type 2 interface module (
) outside the Ex-area. The connection is therefore
established via RS485.