START-UP AND PARAMETER SETTINGS
4
4.1.2
Fitting and connecting the sender/receiver unit on the duct
Connecting the sender/receiver unit to the purge air supply
▸
Check whether the purge air supply is available (the flow direction must be correct and
the purge air hose fits tight on the connection).
▸
Push the DN 40 purge air hose onto the sender/receiver unit connection and secure
with a strap retainer.
Assembly without weather hood
▸
Lay the seal on the flange with tube, position the sender/receiver unit in the flange with
tube and fasten with the assembly kit.
▸
Connect the connection line to the plug connector and screw tight.
Connections for connection line and purge air hose must always be at the bottom
(