Chapter
7
Operating Instructions
CLV640 Bar Code Scanner
70
©
SICK AG · Division Auto Ident · Germany · All rights reserved
8011975/0000/2008-04-14
Startup and configuration
7.5
Default setting
The values of the default setting are permanently saved in the bar code scanner (ROM) and
in the database of the SOPAS-ET configuration software in the device-specific jar file (see
chapter 7.4 First startup, page 68
). A PC is not required to start up the bar code scanner
with the default setting.
7.5.1
Resetting the default setting in the bar code scanner
Prerequisite
The SOPAS-ET configuration software is connected online to the bar code scanner.
Two default setting types can be called up via the SOPAS-ET configuration software:
•
Complete default setting (L
OAD
FACTORY
DEFAULT
)
SOPAS-ET resets all parameter values of the bar code scanner to default. Settings
which have been previously made for the communication parameters of the Ethernet
interfaces or serial data interfaces (e.g. Ethernet address) are overwritten. The connec-
tion(s) to the bar code scanner might be interrupted and has (have) to be reconfigured.
•
Application-specific default setting (L
OAD
APPLICATION
DEFAULT
)
SOPAS-ET resets the parameter values of the bar code scanner but does not change
the communication parameters. Settings which have been previously made for the
communication parameters of the Ethernet interfaces or serial data interfaces are kept
and the current connection(s) to the bar code scanner remain(s) established.
1. In order to discard changes to the parameter set as described above, select the corre-
sponding command in the menu bar under CLV640.
The SOPAS-ET configuration software loads the default setting from the bar code scan-
ner and displays the parameter values in the register tabs. In the bar code scanner, the
default setting will first be active in the temporary main memory only.
The default setting can also be saved on or printed via the PC,
play and print the current parameter set, page 69
2. In the menu bar under T
OOLS
select the command L
OGIN
D
EVICE
.
3. In the dialog window under U
SERLEVEL
in the list box select the entry A
UTORIZED
C
LIENT
.
If the parameter set is password-protected, enter the password "client" in P
ASSWORD
.
4. Click on OK to confirm the dialog window.
5. In the menu bar under CLV640 select the command P
ARAMETER
/S
AVE
P
ERMANENT
.
The SOPAS-ET configuration software transfers the default setting to the permanent pa-
rameter memory (PROM) of the bar code scanner.
If the bar code scanner is equiped with the Micro SD memory card or connected to a
connection module CDB620/CDM420 with parameter memory module CMC600, the
default setting will be permanently transferred to parameter memory CMC600 or mem-
ory card respectively.
Important
Once the default setting has been restored, password-protection is deactivated.