O P E R AT I N G I N S T R U C T I O N S | C F P C U B I C
8021989 / 2017-10-20 | SICK AG
Subject to change without notice
3 6
8
OPERATION
1
2
3
Fig. 6:
CFP Cubic measuring range
1
Level
2
Offset
3
Measuring range
Note:
If the offset parameter is changed, the SPx/RPx/FLx/FHx/QaLow/QaHigh parameters
are automatically adjusted.
8.5.12
Resetting the calibration
Executing this function resets the empty adjustment that was carried out by the cus-
tomer, and the threshold value (
EXPRT / MEAS / TRSHLD
) and the filter time of the
medium adjustment (EXPRT / MEAS / ADAPT) are reset to the default value.
Configuration
1.
Log in to Exprt mode, see
2. Select MEAS / RESET
and confirm using Set.
3.
Confirm the
RESET? safety prompt using Set.