8017035/ZMO9/2017-06-08 • © SICK AG • Subject to change without notice.
127
Serial Communication
Function
Set or
acquire
Send
or
reply
Length Command
(hex)
Data string
Parameter
Calc target2
Set
Send
2
8004
000a000b
a: Calculation
a
Setting value
a
Setting value
4
Calc1
5
Calc2
b: Area
b
Setting value
b
Setting value
0
Area1
2
Area3
1
Area2
3
Area4
Reply
0
—
Acquire Send
1
8005
000a
Reply
1
000b
8.5.7
Output Settings
Function
Set or
acquire
Send
or
reply
Length Command
(hex)
Data string
Parameter
Out target
Set
Send 2
a00e
000a000b
a: Output
a
Setting value
a
Setting value
0
OUT1
2
OUT3
1
OUT2
3
OUTA
b: Output target
b
Setting value
b
Setting value
0
Area1
3
Area4
1
Area2
4
Calc1
2
Area3
5
Calc2
Reply 0
—
Acquire Send 1
a00f
000a
Reply 1
000b
Upper limit
(OUT1 to
OUT3)
20mA
(OUTA)
Set
Send 2
a00a
000abbbb
a: Output
a
Setting value
a
Setting value
0
OUT1
2
OUT3
1
OUT2
3
OUTA
bbbb: Upper limit or 20mA value
Upper limit: -32766 to +32767
20mA: -31767 to +32767
Reply 0
—
Acquire Send 1
a00b
000a
Reply 1
bbbb